Tigo SF

TIGO SF is an accurate and compact CMM specifically designed for use on the shop floor. With a measurement volume of 500 x 580 x 500 mm (X/Y/Z), it is the ideal measurement solution for a wide range of small and medium size parts from all industrial sectors.

Technology Inside

Scan Pilot

Guiding the CMM through the unknown
Scanning of unknown paths can often be challenging and time-consuming. The advanced firmware algorithms of Scan Pilot offer greater motion control capabilities that ensure robust scanning performance.

Fly2 Mode

Gliding smoothly through the measurements
Fly2 Mode further optimises machine motion and minimises idle times so that the machine glides smoothly through its movement trajectories. As a result, program execution times are decreased considerably

Eco Mode

Save energy, sustain the environment
Eco Mode was implemented to reduce the operating cost of CMMs by decreasing energy consumption, and to help protect and sustain the environment. Eco Mode automatically powers down the machine when it stands idle.

XT 40 °C

Extended
temperature

A network of thermal sensors combined with enhanced structural machine temperature compensation, ensures optimum machine performance in harsh environments even at high temperatures and gradients.

TIGO SF specifications

Articulating head with scanning probe HP-S-X1S, Scanning probe heads HP-S- X1C

Max. Permissible Error MPE (µm) and Max. Permissible Limit
MPL (µm) according to ISO 10360-2:2009
Volumetric length measuring error 1)Repeatability range
MPE(E0/E60)MPL(R0)
Temperature range T1 (T1: 18 - 22 °C)2.2 + L/3001.6
Temperature range T2 (T2: 16 - 26 °C)2.5 + L/250
Temperature range T3 (T3: 15 - 30 °C))2.7 + L/200
Temperature range T4 - XT option: (T4: 15 - 40 °C)3.7 + L/100
Temperature range T5 - XT option (T5: 15 - 40 °C)3.7 + L/80
Max. Permissible Error MPE (µm) according to ISO 10360-5:2010
Single stylus form errorMPE(PFTU)2.2
Max. Permissible Error MPE (µm) and Max. Permissible Time MPT (s) according to ISO 10360-4:2000
Single stylus form error, scanning 2) MPE(THP)/MPT(τ)3.5/50

Articulating head with HP-THDe trigger probe

Max. Permissible Error MPE (µm) and Max. Permissible Limit
MPL (µm) according to ISO 10360-2:2009
Volumetric length measuring error 1)Repeatability range
MPE(E0/E60)MPL(R0)
Temperature range T1 (T1: 18 - 22 °C)2.4 + L/3001.8
Temperature range T2 (T2: 16 - 26 °C)2.7 + L/250
Temperature range T3 (T3: 15 - 30 °C))2.9 + L/200
Temperature range T4 - XT option: (T4: 15 - 40 °C)3.9 + L/100
Temperature range T5 - XT option (T5: 15 - 40 °C)3.9 + L/80
Max. Permissible Error MPE (µm)
according to ISO 10360-5:2010
Single stylus form errorMPE(PFTU)2.4

Articulating head with HP-TMe trigger probe

Max. Permissible Error MPE (µm) and Max. Permissible Limit MPL (µm)
according to ISO 10360-2:2009
Volumetric length measuring error 1)Repeatability range
MPE(E0/E60)MPL(R0)
Temperature range T1 (T1: 18 - 22 °C) 2.6 + L/3002.0
Temperature range T2 (T2: 16 - 26 °C)2.9 + L/250
Temperature range T3 (T3: 15 - 30 °C))3.1 + L/200
Temperature range T4 - XT option: (T4: 15 - 40 °C)4.1 + L/100
Temperature range T5 - XT option (T5: 15 - 40 °C)4.1 + L/80
Max. Permissible Error MPE (µm) according to ISO 10360-5:2010
Single stylus form errorMPE(PFTU)2.6
Dynamics
Max. 3D Speed520 mm/s
Max. 3D Acceleration1750 mm/s2
Max. Scanning Speed300 mm/s
Performance test according to the following specificationAmbient temperatureMax. air temperature varMax. gradient in space
Temperature range T118 °C - 22 °C1 °C/h - 2 °C/24h1 °C/m
Temperature range T216 °C - 26 °C1 °C/h - 5 °C/24h
Temperature range T315 °C - 30 °C2 °C/h - 5 °C/24h
Temperature range T4 15 °C - 40 °C2 °C/h - 5 °C/24h
Temperature range T515 °C - 40 °C2 °C/h - 10 °C/24h

Probe heads and sensors

Technical characteristicsHP-S-X1C
Overtravel range± 2 mm in all axes
Stylus jointM3
Max. stylus weight33 g
Max. stylus length Vertical: up to 225 mm
Horizontal: up to 100 mm
Technical characteristicsHH-A-M5 / HH-A-T5
Indexable Probe Head
HH-A-M7.5 / HH-A-T7.5
Indexable Probe Head
Angular rotation A axis: +90° / -115°
B axis: ±180°
A axis: +105° / 0°
B axis: ±180°
Angular rotation step 7.5°
Max. applied torque0.6 Nm
Max. extension length300 mm

Technical characteristics

CharacteristicsSpecifications
Mechanical frameX: T Frame made of welded steel;
Y carriage: made in steel;
Z axis: Micromachined steel extrusion
Surface plateMaterial: Granite
Flatness: according to DIN 876/III
Part Locking: threaded inserts M8 x 1.25e
WeightMax. Part Weight: 150 kg; CMM Weight approx.: 800 kg
Sliding systemDual lineas guide with recirculating ball bearings on all axes
Measuring systemMETALLUR® linear scales. System Resolution: 0.039 μm
Temperature compensationMulti-sensor temperature compensation technology
Ram counterbalanceSteel spring
Supply RequirementsPower. 100/120/230/240 V ± 10% - 50/60 Hz - 2.5 KVA (+0.7 KVA - XT option)
Air. no air required
ConsumptionPower. 0.4 KVAh (+0.55 KVAh - XT option)
Operating SpecificationsAmbient temperature: 10 - 40 °C
Relative humidity: 20% - 90 % non-condensing

Let’s start project together!

Please fill out the form to request a quote. We will get back to you within 24 hours of receiving your request. Thank you for your patience

    What is your name?*

    What company do you represent?*

    Phone number?*

    Tell us about your project or application*